INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-33729
- Event Type
- Malfunction
- Date Received
- October 5, 2014
- Date of Event
- September 5, 2014
- Report Date
- September 5, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
IT WAS REPORTED BY THE CUSTOMER'S GRANDMOTHER THAT THEIR INSULIN PUMP WAS ALARMING NO DELIVERY. CUSTOMER VERIFIED CONNECTION BETWEEN THE RESERVOIR AND INFUSION SET WAS SECURE. CUSTOMER STATED NOT STRAIN AND INFUSION SET IS NOT BENT OR KINKED. DURING TROUBLESHOOTING, CUSTOMER WAS ASKED TO PERFORM A FIXED PRIME OF INSULIN. CUSTOMER STATES THE INSULIN DID EXIT. NEXT, CUSTOMER WAS ASKED TO MANUAL PRIME OF INSULIN. CUSTOMER STATES THE INSULIN DID EXIT AND DEVICE DID NOT ALARM. ADVISED CUSTOMER DEVICE IS FUNCTIONING AS DESIGNED. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 340 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621950 | INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-XXX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR |