FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4142408 · Received October 5, 2014

Report

Report Number
2032227-2014-33784
Event Type
Malfunction
Date Received
October 5, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, COMPROMISED FORCE SENSOR, NO DELIVERY AND MOTOR TESTS. THERE WAS NO MOTOR ERROR ALARM NOTED. THE MOTOR TESTED OUTSIDE OF DEVICE AND PASSED. THE UNIT HAD MINOR SCRATCHES TO LCD WINDOW, CRACKED CASE NEAR LCD WINDOW CORNER, RESERVOIR TUBE LIP, BELT CLIP SLOT, SCRATCHED RESERVOIR TUBE WINDOW, STAINED ADDRESS OR SERIAL NUMBER LABEL AND STAINED END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THE INSULIN PUMP ALARMED MOTOR ERROR. NOTHING FURTHER WAS REPORTED.