FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 4142402 · Received October 5, 2014

Report

Report Number
2032227-2014-33776
Event Type
Malfunction
Date Received
October 5, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP PASSED BASIC OCCLUSION TEST AND DISPLACEMENT TEST. NO MOTOR ERROR ALARMS WERE NOTED. HOWEVER, INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS AND MINOR SCRATCHES ON DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621801 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-715LWWS

Patients

Seq Age Sex Outcome Treatment
1