FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4142326 · Received October 5, 2014

Report

Report Number
2032227-2014-33689
Event Type
Injury
Date Received
October 5, 2014
Date of Event
September 3, 2014
Report Date
October 14, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE FUNCTIONAL TEST, INCLUDING THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST. THE INSULIN PUMP HAD A CRACKED CASE AT A DISPLAY WINDOW CORNER, A CRACKED RESERVOIR TUBE, A CRACKED RESERVOIR TUBE LIP, A CRACKED BELT CLIP SLOT AND MINOR SCRATCHES ON THE DISPLAY WINDOW. NO CRACK WAS NOTED ON THE DISPLAY WINDOW. NO DAMAGE WAS NOTED ON THE DISPLAY GLASS.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN PROVIDED BY FAILURE ANALYSIS THAT WAS NOT INCLUDED ON THE INITIAL DEVICE EVALUATION: THE INSULIN PUMP PASSED THE DELIVERY VOLUME ACCURACY TEST.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP WAS CRACKED ON THE RESERVOIR COMPARTMENT AND DISPLAY SCREEN. CUSTOMER ALSO REPORTED THAT THE PARAMEDICS WERE RECENTLY CALLED TO RESPOND TO A LOS BLOOD GLUCOSE LEVEL OF 32 MG/DL. CUSTOMER REPORTED LOSING CONSCIOUSNESS AT THE TIME OF THIS EVENT, FALLING, AND BREAKING SEVERAL RIBS. BLOOD GLUCOSE LEVEL WAS 110 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621916 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention