FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 4142217 · Received October 5, 2014

Report

Report Number
2939301-2014-26428
Event Type
Malfunction
Date Received
October 5, 2014
Report Date
September 27, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (12/01/2014). THE PATIENT¿S METER HAS BEEN RETURNED ON 10/17/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 11/18/2014 WITH THE FOLLOWING FINDINGS:NO ERROR MESSAGE IS OBSERVED IN THE ERROR LOG, AND ERROR WAS NOT REPRODUCED DURING THE INVESTIGATION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 2 ¿ (01/13/2015). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 12/11/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 1/5/2015 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. WHEN TEST STRIPS WERE TESTED WITH CONTROL SOLUTION, AN ERROR 4 WAS OBSERVED WITH NO ASSIGNABLE CAUSE FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING ERROR UNKNOWN. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
622017 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3683432

Patients

Seq Age Sex Outcome Treatment
1