OT VERIO IQ METER
Report
- Report Number
- 2939301-2014-26422
- Event Type
- Malfunction
- Date Received
- October 5, 2014
- Report Date
- September 27, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (12/04/2014). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 10/28/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 11/24/2014 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. WHEN TEST STRIPS WERE TESTED WITH CONTROL SOLUTION, AN ERROR 4 WAS OBSERVED WITH NO ASSIGNABLE CAUSE FOUND. A SECONDARY ISSUE WAS NOTED WHEN THE TEST STRIPS WERE FOUND TO HAVE RESULTS GREATER THAN +50% OF THE CONTROL SOLUTION WHEN TESTED WITH CONTROL SOLUTION. IN ADDITION, A TERTIARY ISSUE WAS NOTED WHEN THE TEST STRIPS WERE FOUND TO HAVE RESULTS ABOVE RANGE WHEN TESTED WITH CONTROL SOLUTION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING ERROR 4. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 622011 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3643782 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |