FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4142124 · Received October 4, 2014

Report

Report Number
2032227-2014-33152
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
August 29, 2014
Report Date
September 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A BUTTON ERROR DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. THE INSULIN PUMP WAS PROGRAMMED WITH MULTIPLE BOLUS DELIVERIES AND ALL REGISTERED PROPERLY IN THE BOLUS HISTORY. NO BOLUS HISTORY ANOMALY WAS NOTED. NO RESET ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP WAS RECEIVED WITH A PEELING, BUT NOT WORN OR FADED, ADDRESS AND SERIAL NUMBER LABEL, MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED RESERVOIR TUBE LIP AND A MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP AND THE BOLUS HISTORY WAS RECORDING WRONG FOR ABOUT A MONTH. IT WAS NOTED THAT THE DATE ON THE INSULIN PUMP WAS WRONG. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 82 MG/DL AND HAS TREATED WITH A BOLUS. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621121 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722RNAS

Patients

Seq Age Sex Outcome Treatment
1 75 YR