FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4142113 · Received October 4, 2014

Report

Report Number
2032227-2014-33163
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD UNRESPONSIVE ESCAPE, ACT, DOWN AND UP BUTTONS DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT A DISPLAY WINDOW CORNER, A CRACKED RESERVOIR TUBE LIP AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED A BUTTON WAS UNRESPONSIVE ON THE CUSTOMER'S INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS NOT KNOWN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621212 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-754WWB

Patients

Seq Age Sex Outcome Treatment
1