PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-33161
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Date of Event
- September 5, 2014
- Report Date
- September 5, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. THE INSULIN PUMP WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM ON THE INSULIN PUMP, AFTER IT MAY HAVE BEEN EXPOSED TO PERSPIRATION. THE CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THE ALARM. HER BLOOD GLUCOSE LEVEL WAS 113 MG/DL. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP TREATMENT PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621348 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |