530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-33357
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Date of Event
- August 29, 2014
- Report Date
- September 5, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED LCD WINDOW, CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).
CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A MOTOR ERROR ALARM DURING BOLUS. CUSTOMER DID NOT REPORT ANY SPECIFIC EVENTS LEADING UP TO THE ERROR ALARM. BLOOD GLUCOSE LEVEL AT THE TIME OF THE ERROR ALARM WAS NOT PROVIDED, BUT CUSTOMER REPORTED THAT IT WAS HIGH. BLOOD GLUCOSE LEVEL AT THE TIME OF THE CALL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621299 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC MINIMED | MMT-551NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |