FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4142077 · Received October 4, 2014

Report

Report Number
2032227-2014-33333
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE UNIT PASSED THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME, COMPROMISED FORCE SENSOR TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. THERE WAS NO MOTOR ERROR ALARMS OR DISPLACEMENT ANOMALIES NOTED. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. THE UNIT WAS RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND RESERVOIR TUBE LIP. THE UNIT WAS RECEIVED WITH MINOR SCRATCHED LCD WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 173MG/DL. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621106 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 53 YR