FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4142033 · Received October 4, 2014

Report

Report Number
2032227-2014-33325
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNIT RECEIVED WITH CRACKED CASE ON DISPLAY WINDOW CORNERS, MINOR SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP, AND CRACKED BATTERY TUBE THREADS. INTERMITTENT BUTTON RESPONSE NOTED DURING TESTING DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.(B)(4).

Description of Event or Problem · 1

IT IS REPORTED THAT CUSTOMER EXPERIENCED PHYSICAL DAMAGE TO INSULIN PUMP. IT IS REPORTED THAT CASE IS BROKEN AT BATTERY COMPARTMENT. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621750 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWL

Patients

Seq Age Sex Outcome Treatment
1