FDA Adverse Event
Malfunction
Summary report: N
PARADIGM INSULIN INFUSION PUMP
MDR report key: 4142002
·
Received October 4, 2014
Report
- Report Number
- 2032227-2014-33135
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Date of Event
- September 5, 2014
- Report Date
- September 5, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LZG
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A LOOSE AND PROTRUDED DRIVE SUPPORT DISK. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED RESERVOIR TUBE LIP.
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED AN ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS 130 MG/DL. THE CUSTOMER WAS ADVISED TO REVERT TO A BACKUP PLAN BECAUSE THE INSULIN PUMP NEEDED TO BE REPLACED. NOTHING FURTHER WAS REPORTED.