FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141981 · Received October 4, 2014

Report

Report Number
2032227-2014-33119
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED AN ERROR AND INSULIN WAS SQUIRTING OUT. . THE CUSTOMER STATED THAT WHILE THEY WERE ATTEMPTING PERFORM THE PRIMING PROCESS, THE NUMBERS ON THE INSULIN PUMP WOULD NOT COME UP. THE DRIVE SUPPORT CAP WAS RECESSED. THE CUSTOMER'S BLOOD GLUCOSE WAS 208 MG/DL. THE CUSTOMER STATED THAT THEY WOULD GET A BACKUP PLAN THE DAY AFTER THE PHONE CALL. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. THE CUSTOMER ALSO STATED THAT THEY HAD THE INSULIN PUMP ATTACHED TO THEM WHILE THEY HAD A CAT SCAN PROCEDURE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621767 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 58 YR