PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-33118
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Date of Event
- September 5, 2014
- Report Date
- September 5, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
NO BUTTON ERROR ALARM WAS NOTED. THE UP ARROW BUTTON DID NOT RESPOND DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, A BROKEN BELT CLIP SLOT AND A CRACKED RESERVOIR TUBE LIP.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. WHEN THE CUSTOMER ATTEMPTED DELIVERING A BOLUS, THE NUMBERS STARTED GOING UP ON THEIR OWN. THE CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621739 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |