FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141953 · Received October 4, 2014

Report

Report Number
2032227-2014-32933
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 4, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO UNEXPECTED KEYPAD ANOMALIES WERE NOTED. NO UNEXPECTED BUTTON ERROR ALARMS WERE NOTED AND THE BUTTONS ALL FUNCTIONED PROPERLY. HOWEVER, CORRODED KEYPAD TRACES WERE NOTED DURING THE VISUAL INSPECTION. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A BROKEN RESERVOIR TUBE LIP, SCRATCHES ON THE RESERVOIR TUBE WINDOW, CRACKED RESERVOIR TUBE, CRACKED BATTERY TUBE THREADS AND A BROKEN BELT CLIP SLOT.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD AN UNRESPONSIVE KEYPAD AND BUTTONS; THE BUTTONS WERE NOT RELEASING. LATER, DURING THE DAY, PRIOR TO THE PHONE CALL THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS 278 MG/DL. THE CUSTOMER STATED THAT THE REASON FOR THE HIGH BLOOD GLUCOSE WAS THAT THEY HAVE NOT HAD INSULIN FOR A COUPLE OF HOURS DUE THE ISSUED WITH THE DEVICE. THE CUSTOMER STATED THAT THE INSULIN PUMP MIGHT HAVE BEEN EXPOSED TO MOISTURE OR PERSPIRATION DUE TO HUMIDITY. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621528 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAH

Patients

Seq Age Sex Outcome Treatment
1 42 YR