FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141917 · Received October 4, 2014

Report

Report Number
2032227-2014-33178
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE, DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE AND MISSING END CAP STICKER. INSULIN PUMP RECEIVED WITH CURRENTS IN SPECIFICATIONS. NO WEAK BATTERY.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING BUTTON ERROR. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LED TO THE ALARM OR IF THE DEVICE WAS DROPPED. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 184 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621517 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAS

Patients

Seq Age Sex Outcome Treatment
1 27 YR