FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 4141909 · Received October 4, 2014

Report

Report Number
2032227-2014-32913
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 4, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LZG
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO UNEXPECTED BUTTON ERROR ALARMS NOTED. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE ON THE ESC BUTTON DUE TO CORRODED KEYPAD TRACES. THE DEVICE HAD CRACKED LCD WINDOW, CRACKED CASE AT LCD WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, SCRATCHES ON RESERVOIR TUBE WINDOW, AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR AND THE CUSTOMER WAS UNABLE TO CLEAR IT. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621736 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-715WWL

Patients

Seq Age Sex Outcome Treatment
1