FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 4141837 · Received October 4, 2014

Report

Report Number
1416980-2014-34524
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 8, 2014
Report Date
September 10, 2014
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K071222
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS LOT WAS MANUFACTURED FROM JUNE 26, 2014 TO JUNE 27, 2014. EVALUATION SUMMARY: THE SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION IDENTIFIED A SINGLE BLACK IRREGULARLY SHAPED PARTICLE, APPROXIMATELY 0.02 MM IN SIZE, EMBEDDED IN A STRESS-MEMBER PLUG MATERIAL. THE CAUSE OF THE PRESENCE OF THE PARTICULATE MATTER COULD NOT BE DETERMINED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A SMALL BLACK PARTICLE ¿NEAR THE END OF THE CENTRAL STEM INSIDE THE BALLOON¿ OF A LARGE VOLUME INFUSOR. THIS WAS NOTED DURING FILLING OF THE DEVICE WITH 5-FLUOROURACIL AND SODIUM CHLORIDE 0.9%. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621472 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 14F054