FDA Adverse Event Malfunction Summary report: N

FOLFUSOR

MDR report key: 4141834 · Received October 4, 2014

Report

Report Number
1416980-2014-34521
Event Type
Malfunction
Date Received
October 4, 2014
Report Date
September 10, 2014
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). VISUAL INSPECTION NOTED FLUID IN THE BAG THAT CONTAINED THE DEVICE. THE LEAK WAS DISCOVERED TO BE COMING FROM AN UNTIGHTENED BLUE WINGED LUER CAP. A FUNCTIONAL LEAK TEST WAS PERFORMED BY FILLING THE DEVICE WITH WATER AND TIGHTENING THE BLUE WINGED LUER CAP. THE NEXT DAY THERE WERE NO SIGNS OF A LEAK FROM THE LUER CAP. THE REPORTED CONDITION WAS NOT VERIFIED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. UPON COMPLETION OF THE INVESTIGATION, OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LARGE VOLUME FOLFUSOR HAD LEAKED FLUOROURACIL INTO THE OVERPOUCH. THIS WAS NOTED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 1 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621752 FOLFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 14G012

Patients

Seq Age Sex Outcome Treatment
1