Report
- Report Number
- 1416980-2014-34521
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Report Date
- September 10, 2014
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). VISUAL INSPECTION NOTED FLUID IN THE BAG THAT CONTAINED THE DEVICE. THE LEAK WAS DISCOVERED TO BE COMING FROM AN UNTIGHTENED BLUE WINGED LUER CAP. A FUNCTIONAL LEAK TEST WAS PERFORMED BY FILLING THE DEVICE WITH WATER AND TIGHTENING THE BLUE WINGED LUER CAP. THE NEXT DAY THERE WERE NO SIGNS OF A LEAK FROM THE LUER CAP. THE REPORTED CONDITION WAS NOT VERIFIED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). (B)(6). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. UPON COMPLETION OF THE INVESTIGATION, OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A LARGE VOLUME FOLFUSOR HAD LEAKED FLUOROURACIL INTO THE OVERPOUCH. THIS WAS NOTED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 1 OF 2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621752 | FOLFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 14G012 |