FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4141793 · Received October 4, 2014

Report

Report Number
2032227-2014-32976
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 4, 2014
Report Date
March 19, 2019
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM NOTED DURING TESTING. DEVICE HAD UNRESPONSIVE ESC BUTTON DUE TO CORRODED KEYPAD TRACES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM ON HIS INSULIN PUMP, AFTER HE SLEPT WITH IT UNDER HIS PILLOW. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN. HIS BLOOD GLUCOSE LEVEL WAS 303 MG/DL, WHICH HE TREATED WITH AN INSULIN INJECTION. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621436 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAS

Patients

Seq Age Sex Outcome Treatment
1 20 YR