FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141747 · Received October 4, 2014

Report

Report Number
2032227-2014-33521
Event Type
Injury
Date Received
October 4, 2014
Date of Event
September 2, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT, RESTART ERROR TEST, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY ALARM TESTS. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP AND CRACKED DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER'S MOTHER REPORTED HER DAUGHTER HAD A HIGH BLOOD GLUCOSE OF 574 MG/DL. THE DAUGHTER TREATED HER HIGH BLOOD GLUCOSE WITH THE INSULIN PUMP. THE MOTHER ALSO REPORTED SEVERAL CRACKS ON THE FRONT OF HER DAUGHTER'S INSULIN PUMP. THE MOTHER WAS UNSURE HOW THE DAMAGE HAD OCCURRED. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621063 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 19 YR