PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4141738 · Received October 4, 2014

Report

Report Number
2032227-2014-32905
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 4, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD AN UNRESPONSIVE ESCAPE BUTTON DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THERE IS A BUTTON ERROR ALARM. THE BLOOD GLUCOSE READING IS UNKNOWN. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.