PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-33504
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Date of Event
- September 4, 2014
- Report Date
- September 4, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
THE INSULIN PUMP PASSED THE REWIND, DISPLACEMENT, PRIME AND EXCESSIVE NO DELIVERY ALARM TEST. NO EXCESSIVE NO DELIVERY ALARM WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WIDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED NO DELIVERY AND THAT INSULIN WAS "SQUIRTING" OUT DURING THE MANUAL PRIME PROCESS. THE BLOOD GLUCOSE READING WAS 382 MG/DL, WHICH THE CUSTOMER TREATED WITH A MANUAL INJECTION. THE CALLER STATED THAT THE INSULIN PUMP DID NOT DETECT THE SENSOR AND THAT IT KEPT PUSHING INSULIN OUT AFTER AN INFUSION SET CHANGE. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621167 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |