FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4141722 · Received October 4, 2014

Report

Report Number
2032227-2014-32897
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 4, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS, AND A CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP'S UP ARROW BUTTON WAS RESPONDING INTERMITTENTLY OR NOT AT ALL. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621156 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWS

Patients

Seq Age Sex Outcome Treatment
1