PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-33417
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Date of Event
- September 5, 2014
- Report Date
- September 5, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP PASSED BASIC OCCLUSION TEST AND DISPLACEMENT TEST. NO MOTOR ERROR ALARMS WERE NOTED. HOWEVER, THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR. THE MOTOR WAS TESTED OUTSIDE OF THE INSULIN PUMP AND PASSED. THE INSULIN PUMP DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH CRACKED CASE ON THE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON DISPLAY WINDOW.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM ON THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS NOT KNOWN. THE CUSTOMER REPORTEDLY DID NOT KNOW IF THE DRIVE SUPPORT CAP WAS LOOSE. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621015 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722WWS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |