FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141681 · Received October 4, 2014

Report

Report Number
2032227-2014-33298
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 4, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED FUNCTIONAL TESTS. NO MOTOR ERROR ALARM NOTED. THE PRIME TEST COULD NOT BE COMPLETED DUE TO FAULTY FORCE SENSOR RELAY. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED LIQUID CRYSTAL DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP, CRACKED BELT CLIP SLOT, AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER REPORTS TO HAVE RECEIVED A MOTOR ERROR ALARM DURING PRIMING. CUSTOMER'S BLOOD GLUCOSE WAS 145 MG/DL. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621366 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 69 YR