FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141642 · Received October 4, 2014

Report

Report Number
2032227-2014-33235
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 3, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME ALARM TEST DUE TO MOISTURE DAMAGE NOTED IN FORCE SENSOR. INSULIN PUMP PASSED BASIC OCCLUSION AND DISPLACEMENT TEST. INSULIN PUMP HAD AN INTERMITTENT MOTOR ERROR ALARM DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH. INSULIN PUMP HAD A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW, MISSING END CAP STICKER AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING INSULIN BOLUS DELIVERY. CUSTOMER RECEIVED A SECOND MOTOR ERROR ALARM THE NEXT DAY AFTER BOLUS DELIVERY. CUSTOMER STATED THAT THE DEVICE HAS FALLEN OUT OF HER POCKET A COUPLE OF TIMES. DEVICE WAS EXPOSED TO MAGNETIC FIELDS; CUSTOMER USES A MAGNETIC SENSOR AT WORK TO REMOVE CLOTHING DEVICE. CUSTOMER DOES NOT USE THE SENSOR FEATURE AND IS ABLE TO REWIND THE INSULIN PUMP. BLOOD GLUCOSE LEVEL IS 538 MG/DL; TREATED WITH BOLUS DOSE. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621062 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 18 YR