FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141614 · Received October 4, 2014

Report

Report Number
2032227-2014-33206
Event Type
Injury
Date Received
October 4, 2014
Date of Event
August 27, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
RI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INSULIN PUMP PASSED FUNCTIONAL TEST INCLUDING PRIME, DISPLACEMENT, BASIC OCCLUSION, OCCLUSION AND EXCESSIVE NO DELIVERY ALARM TESTS. CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED RESERVOIR TUBE, BROKEN BELT CLIP SLOT AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE. BLOOD GLUCOSE VALUE AT THE TIME OF THE ADMISSION WAS 600 MG/DL; TREATED WITH INSULIN DRIP. DURING TROUBLESHOOT; IT WAS STATED THE DRIVE SUPPORT CAP IS RECESSED. CUSTOMER STATED THE TUBING HAS NO AIR BUBBLES. CUSTOMER WAS ABLE TO REWIND AND PERFORM MANUAL PRIME; INSULIN DID EXIT THE TUBING. CUSTOMER DID NOT FIND AN INSULIN LEAK. CUSTOMER STATED THAT TIME, DATE, BASAL RATES AND BOLUS WIZARD ARE SET UP CORRECTLY. CUSTOMER REMOVED THE INFUSION SET AND THE CANNULA IS NOT BENT OR OCCLUDED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621254 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 40 YR Hospitalization