FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4141582 · Received October 4, 2014

Report

Report Number
2032227-2014-32884
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 3, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD UNRESPONSIVE BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM ON THE INSULIN PUMP. HE ALSO REPORTED THAT THE INSULIN PUMP WAS UNRESPONSIVE. THE BLOOD GLUCOSE READING WAS 150 MG/DL. HE STATED THAT THE INSULIN PUMP MAY HAVE BEEN EXPOSED TO SWEAT. CUSTOMER REVERTED TO TREATING WITH AN INSULIN PEN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621334 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 44 YR