FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141560 · Received October 4, 2014

Report

Report Number
2032227-2014-32850
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
September 4, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT TEST. THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A LOOSE AND PROTRUDED DRIVE SUPPORT DISK. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT A DISPLAY WINDOW CORNER, A CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE WINDOW, CRACKED BELT CLIP SLOT, STAINED END CAP STICKER AND A STAINED ADDRESS AND SERIAL NUMBER LABEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM AS WELL AS A COMPROMISED FORCE SENSOR ALARM. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621314 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 31 YR