FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141558 · Received October 4, 2014

Report

Report Number
2032227-2014-32847
Event Type
Malfunction
Date Received
October 4, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM NOTED. ESC BUTTON DID NOT RESPOND DUE TO CORRODED KEYPAD TRACE. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP, MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 243MG/DL. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621239 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 44 YR