PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-32873
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Date of Event
- September 3, 2014
- Report Date
- September 4, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW AND CRACKED RESERVOIR TUBE LIP.
THIS MDR RELATED TO THE (B)(4)MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC (B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED THAT THE ACT BUTTON ON THE INSULIN PUMP WAS NOT FULLY RESPONSIVE. HE STATED THAT THE INSULIN PUMP HAD BEEN EXPOSED TO SWEAT. HE DID NOT RECALL THE BLOOD GLUCOSE READING. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621237 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |