OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 3008011247-2014-00080
- Event Type
- Malfunction
- Date Received
- October 4, 2014
- Date of Event
- September 12, 2014
- Report Date
- September 15, 2014
- Manufacturer
- TRIVASCULAR, INC.
- Product Code
- MIH
- PMA / PMN Number
- P120006
- Removal / Correction Number
- REPORTED - TBA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
THIS FOLLOW-UP REPORT IS TO CORRECT INFORMATION SUBMITTED IN THE ORIGINAL REPORT WHICH STATED THAT THE PATIENT EXPERIENCED HYPOTENSION. THE PATIENT DID NOT EXPERIENCE HYPOTENSION DURING THE IMPLANT PROCEDURE.
BASED UPON THIS EVENT AND AN INCREASE IN COMPLAINTS RELATED TO INCOMPLETE INTRAOPERATIVE POLYMER FILL OF THE OVATION PRIME 29MM AORTIC BODY STENT GRAFT, TRIVASCULAR INITIATED A FIELD SAFETY CORRECTIVE ACTION ON SEPTEMBER 20, 2014 TO RETRIEVE FROM DISTRIBUTOR INVENTORY (B)(4) AFFECTED DEVICES MANUFACTURED WITHIN THE SAME TIME PERIOD AS THE DEVICE FOR THIS EVENT. NO DEVICES IN THE UNITED STATES WERE SUBJECT TO THIS CORRECTIVE ACTION.
A 29MM OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT DURING ENDOVASCULAR REPAIR OF AN ABDOMINAL AORTIC ANEURYSM. DURING POLYMER FILL, THE FILL POLYMER SYRINGE EMPTIED AND THE PATIENT EXPERIENCED TRANSIENT HYPOTENSION WHICH WAS SUCCESSFULLY TREATED IN ACCORDANCE WITH THE PRODUCT IFU. THE PHYSICIAN EXPERIENCED DIFFICULTIES CANNULATING THE CONTRALATERAL GATE, WHICH RESULTED IN PROLONGING THE IMPLANT PROCEDURE. A BRACHIAL APPROACH WAS SUCCESSFULLY USED TO CANNULATE THE GATE, DEPLOY THE ILIAC LIMB, AND EXCLUDE THE ANEURYSM WITH NO ADDITIONAL PATIENT SEQUELAE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621174 | OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM | ENDOVASCULAR STENT GRAFT, ABDOMINAL AORTIC ANEURYSM | MIH | TRIVASCULAR, INC. | TV-AB2980-D | FS031814-17 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Required Intervention |