FDA Adverse Event Malfunction Summary report: N

ABBOTT M2000SP

MDR report key: 4141356 · Received October 3, 2014

Report

Report Number
3005248192-2014-00060
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
September 10, 2014
Report Date
September 10, 2014
Manufacturer
ABBOTT MOLECULAR, INC.
Product Code
JJH
PMA / PMN Number
K092705
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
SERVICE PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THE ELEVATED COMPLAINT INVESTIGATION (ECINV) FOR 4210-ECINV DETERMINED THIS COMPLAINT IS THE SAME ISSUE AS THAT OF 3853-ECINV. SUMMARY OF ELEVATED COMPLAINT INVESTIGATION 3853-ECINV: PROBLEM: EXCESSIVE NOISE FROM THE LIHA AND ROMA X-DRIVE MOTORS WAS REPORTED. NOISE AND STEP-LOSS FAILURES OCCUR WITHIN 2-3 MONTHS OF INSTRUMENT INSTALLATION OR REPLACEMENT SPARE PART INSTALLATION. CONCLUSIONS: RETAIN TESTING OF THE X-DRIVE MOTOR WAS NOT APPLICABLE AS NO FILE SAMPLES ARE RETAINED. SUPPLIER TECAN HAS CONFIRMED A PRODUCT DEFICIENCY WITH THE DRIVE BELT PRODUCTION PROCESS ON SOME BELTS. WITHIN 2-3 MONTHS OF OPERATION THE BELT TENSION CHANGES LEADING TO EXCESSIVE NOISE AND IN SOME CASES THE ASSOCIATED ARM CRASHES INTO A WORKTABLE OBJECT OR BECOMES MISALIGNED. X-DRIVE AXIS MOTOR REPLACEMENT SPARE PARTS AND ALL M2000SP AND M24SP INSTRUMENTS MANUFACTURED WITH THE ASSEMBLIES WITH DATE CODES BETWEEN 01 JANUARY 2013 THROUGH 30 SEPTEMBER 2013 ARE SUSPECT. THE M2000SP SERVICE MANUAL CONTAINS ADEQUATE INSTRUCTIONS FOR TROUBLESHOOTING X-DRIVE MOTOR ISSUES WITH REPLACEMENT INSTRUCTIONS AND PART NUMBER INFORMATION. VALID MATERIAL SPECIFICATIONS ARE RESIDENT IN PILGRIM SMARTSOLVE FOR THE MOTOR DC ARM X-AXIS 1000 INCREMENTS EVO PART NUMBER 10649045. A THOROUGH COMPLAINT SEARCH WAS PERFORMED TO IDENTIFY PAST COMPLAINT TICKETS FROM THE LAST TWO YEARS FOR ANY INSTANCES OF THE X-DRIVE MOTORS MAKING EXCESSIVE NOISE ON THE M2000SP INSTRUMENT. COMPLAINT HISTORY REVIEW SHOWED THAT, OVER THE LAST TWO YEARS, THIRTEEN COMPLAINTS INCLUDING THE SUSPECT COMPLAINTS ((B)(4)) WERE FOUND TO BE RELATED TO THE ISSUE OF THE X-DRIVE MOTOR FAILING ON THE M2000SP. (B)(4). ELEVATED COMPLAINT INVESTIGATION 3853-ECINV CONCLUDED THERE WAS A CONFIRMED PRODUCT DEFICIENCY WITH THE X-DRIVE MOTOR PART NUMBER 10649045. ABBOTT MOLECULAR DETERMINED A FIELD ACTION (CORRECTION) SHOULD BE TAKEN TO INSPECT AND, IF NECESSARY, REPLACE THE AFFECTED BELTS. ABBOTT MOLECULAR INITIATED FIELD ACTION NUMBER FA-CAM-NOV2013-177 ON NOVEMBER 23, 2013 AS A RECORDABLE FIELD ACTION. UPON RECEIPT OF FDA 483 OBSERVATION #1 ON JULY 11, 2014, ABBOTT MOLECULAR DETERMINED THIS FIELD ACTION SHOULD HAVE BEEN REPORTED TO FDA PER 21 CFR 806, AND ON JULY 29, 2014 REPORTED IT TO THE FDA (B)(4) DISRICT RECALL COORDINATOR. THEREFORE, THE PURPOSE OF SUBMITTING MDR 3005248192-2014-00060 IS TO RETROSPECTIVELY REPORT THIS COMPLAINT BECAUSE IT IS ASSOCIATED WITH REPORTABLE FIELD ACTION FA-CAM-NOV2013-177 (806 REPORT NUMBER 3005248192-07/28/2014-002-C). TO DATE, AN FDA RECALL NUMBER HAS NOT BEEN RECEIVED. THE FOLLOWING MDRS WERE ALSO REPORTED: MDR 3005248192-2014-00007; MDR 3005248192-2014-00008; MDR 3005248192-2014-00009; MDR 3005248192-2014-00010; MDR 3005248192-2014-00011; MDR 3005248192-2014-00012; MDR 3005248192-2014-00013; MDR 3005248192-2014-00014; MDR 3005248192-2014-00015; MDR 3005248192-2014-00016; MDR 3005248192-2014-00017; MDR 3005248192-2014-00018; MDR 3005248192-2014-00019; MDR 3005248192-2014-00020; MDR 3005248192-2014-00021; MDR 3005248192-2014-00022; MDR 3005248192-2014-00023; MDR 3005248192-2014-00024; MDR 3005248192-2014-00025; MDR 3005248192-2014-00026; MDR 3005248192-2014-00027; MDR 3005248192-2014-00028; MDR 3005248192-2014-00029; MDR 3005248192-2014-00030; MDR 3005248192-2014-00031; MDR 3005248192-2014-00032; MDR 3005248192-2014-00033; MDR 3005248192-2014-00034; MDR 3005248192-2014-00035; MDR 3005248192-2014-00036; MDR 3005248192-2014-00037; MDR 3005248192-2014-00038; MDR 3005248192-2014-00039; MDR 3005248192-2014-00040; MDR 3005248192-2014-00041; MDR 3005248192-2014-00042; MDR 3005248192-2014-00043; MDR 3005248192-2014-00044; MDR 3005248192-2014-00045; MDR 3005248192-2014-00046; MDR 3005248192-2014-00047; MDR 3005248192-2014-00048; MDR 3005248192-2014-00049; MDR 3005248192-2014-00050; MDR 3005248192-2014-00051; MDR 3005248192-2014-00052; MDR 3005248192-2014-00053; MDR 3005248192-2014-00054; MDR 3005248192-2014-00055; MDR 3005248192-2014-00058; MDR 3005248192-2014-00059.

Description of Event or Problem · 1

THE ABBOTT M2000 SYSTEM IS INTENDED FOR USE IN PERFORMING NUCLEIC ACID AMPLIFICATION TESTING BY POLYMERASE CHAIN REACTION IN CLINICAL LABORATORIES. THE SYSTEM IS COMPOSED OF THE M2000RT AND THE M2000SP INSTRUMENTS. THE M2000SP IS INTENDED FOR AUTOMATED SAMPLE PREPARATION AND NUCLEIC ACID EXTRACTION PRIOR TO TESTING. AN ABBOTT FIELD SERVICE ENGINEER (FSE) PERFORMED TECHNICAL SERVICE BULLETIN (TSB) 606-041 "M2000SP X-DRIVE ASSEMBLY BELT INSPECTION AND REPLACEMENT PROCEDURE" AND OBSERVED THAT BOTH X-BELTS FOR LIHA AND ROMA EXHIBITED SIGNS OF CORROSION AND WERE REPLACED. THE INSTRUMENT WAS INACTIVE. THE ACTIVITY WAS PERFORMED AT AN ABBOTT WAREHOUSE. THERE WAS NOT ANY REPORT OF HARM. ELEVATED COMPLAINT INVESTIGATION 4210-ECINV CONFIRMED THAT THIS COMPLAINT IS LINKED TO A KNOWN PRODUCT MALFUNCTION OF THE X-DRIVE MOTOR. (B)(4). MEDICAL EVALUATION CONCLUDED THIS ISSUE IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY. THIS COMPLAINT IS REPORTED IN ACCORDANCE WITH 21 CFR 803 ON THE BASIS IT IS LINKED TO REPORTABLE FIELD ACTION FA-CAM-NOV2013-177 PER 21 CFR 806.