PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141336 · Received October 3, 2014

Report

Report Number
3004209178-2014-89244
Date Received
October 3, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTORNIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNK WHETHER OT NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARMED DURING BASIC OCCLUSION TEST AND PRIME ALARMED DURING PRIME ALARM TEST DUE TO FAULTY FORCE SENSOR. INSULIN PUMP RECEIVED WITH A BROKEN RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, CRACKED DISPLAY WINDOW, CRACKED RESERVOIR WINDOW, CRACKED RESERVOIR TUBE AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION. UNABLE TO CONFIRM EXCESSIVE NO DELIVERY ALARMS DUE TO PRIME ALARMS. MOTOR TESTED OK.

Description of Event or Problem · 1

IT WAS REPORTED THAT CUSTOMER RECEIVED A NO DELIVERY ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 396MG/DL. ADVISED CUSTOMER INSULIN PUMP WILL BE REPLACED. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620932 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTORNIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 48 YR