PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141289 · Received October 3, 2014

Report

Report Number
3004209178-2014-92087
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
July 10, 2014
Report Date
July 10, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM. INSULIN PUMP RECEIVED WITH CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE AND CRACKED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED A BUTTON ERROR ON HER INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 240 MG/DL. INSULIN PUMP WILL BE PLACED. NO ADDITIONAL INFO PROVIDED.