FDA Adverse Event Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4141288 · Received October 3, 2014

Report

Report Number
2032227-2014-91303
Date Received
October 3, 2014
Date of Event
July 9, 2014
Report Date
July 9, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4) SITE, PER VARIANCE 5.

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE ON ELECTRONICS NOTED. INSULIN PUMP RECEIVED WITH SEVERE SCRATCHES ON DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PATIENT HAS A BUTTON ERROR ON THEIR INSULIN PUMP AND THE BUTTON WILL NOT WORK. PATIENT'S BLOOD GLUCOSE LEVEL IS AT 170MG/DL. TROUBLE SHOOTING IS CONDUCTED AND FOUND WATER DAMAGE. DEVICE IS BEING RETURNED. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
619948 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 25 YR