FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4141173 · Received October 3, 2014

Report

Report Number
2032227-2014-32792
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
September 1, 2014
Report Date
September 4, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE PUMP HAD A CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE LCD WINDOW, AND SCRATCHED RESERVOIR TUBE WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING MULTIPLE MOTOR ERROR ALARMS ON THE INSULIN PUMP DURING BASAL. CUSTOMER DOES USE THE SENSOR FEATURE ON THE INSULIN PUMP. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
617773 530G INSULIN PUMP OZO OZO MEDTRONIC MINIMED MMT-751NAL

Patients

Seq Age Sex Outcome Treatment
1 66 YR