SHEATH, 2.9MM OPERATIVE INNER
Report
- Report Number
- 0002936485-2014-00783
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- August 12, 2014
- Report Date
- August 12, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- HIH
- PMA / PMN Number
- K040390
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. VISUAL INSPECTION : THE SHEATH, 2.9MM OPERATIVE INNER WAS RECEIVED WITH A DAMAGED DISTAL TIP AND A PIECE OF METAL WAS MISSING ON THE DISTAL TIP.PROBABLE ROOT CAUSE COULD BE THE SHEATH GOT IN CONTACT WITH A SHARP OBJECT DURING USE OR DURING CLEANING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
UPON RECEIPT OF DEVICE, A NEW FAILURE MODE OF TIP BREAK HAS BEEN IDENTIFIED, THUS CONSTITUTING A NEW AWARENESS DATE OF THE DATE THE INVESTIGATION RESULTS WERE RECEIVED BY STRYKER. THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. VISUAL INSPECTION : THE SHEATH, 2.9MM OPERATIVE INNER WAS RECEIVED WITH A DAMAGED DISTAL TIP AND A PIECE OF METAL WAS MISSING ON THE DISTAL TIP. PROBABLE ROOT CAUSE COULD BE THE SHEATH GOT IN CONTACT WITH A SHARP OBJECT DURING USE OR DURING CLEANING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
IT WAS REPORTED THAT THE TIP OF THE UNIT IS DAMAGED. UPON RECEIPT OF THE PRODUCT, IT WAS DETERMINED THAT THERE ARE PIECES OF METAL MISSING FROM THE DISTAL TIP OF THE DEVICE.
IT WAS REPORTED THAT THE TIP OF THE UNIT IS DAMAGED. UPON RECEIPT OF THE PRODUCT, IT WAS DETERMINED THAT THERE ARE PIECES OF METAL MISSING FROM THE DISTAL TIP OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 619130 | SHEATH, 2.9MM OPERATIVE INNER | HYSTEROSCOPE (AND ACCESSORIES) | HIH | STRYKER ENDOSCOPY-SAN JOSE | 1243848 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |