XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2024168-2014-06429
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- September 7, 2014
- Report Date
- September 8, 2014
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NIQ
- PMA / PMN Number
- P110019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE STENT DISLODGEMENT WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY/REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
IT WAS REPORTED THAT DURING PREPARATION OF A 3.5 X 18 MM XIENCE XPEDITION STENT DELIVERY SYSTEM WHEN REMOVING THE PROTECTIVE SHEATH AND STYLET, THE STENT DISLODGED FROM THE BALLOON. THERE WAS NO RESISTANCE REMOVING THE PROTECTIVE SHEATH. A NEW XIENCE XPEDITION WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 618447 | XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM | DRUG ELUTING CORONARY STENT SYSTEM | NIQ | AV-TEMECULA-CT | 3120341 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |