OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 3008011247-2014-00076
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- September 4, 2014
- Report Date
- September 4, 2014
- Manufacturer
- TRIVASCULAR, INC.
- Product Code
- MIH
- PMA / PMN Number
- P120006
- Removal / Correction Number
- REPORTED - TBA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
BASED UPON THIS EVENT AND AN INCREASE IN COMPLAINTS RELATED TO INCOMPLETE INTRAOPERATIVE POLYMER FILL OF THE OVATION PRIME 29MM AORTIC BODY STENT GRAFT, TRIVASCULAR INITIATED A FIELD SAFETY CORRECTIVE ACTION ON SEPTEMBER 20, 2014 TO RETRIEVE FROM DISTRIBUTOR INVENTORY (B)(4) AFFECTED DEVICES MANUFACTURED WITHIN THE SAME TIME PERIOD AS THE DEVICE FOR THIS EVENT. NO DEVICES IN THE UNITED STATES WERE SUBJECT TO THIS CORRECTIVE ACTION.
A 29MM OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT DURING ENDOVASCULAR REPAIR OF AN ABDOMINAL AORTIC ANEURYSM. THE MID-CROWN OF THE PROXIMAL STENT WAS INADVERTENTLY DEPLOYED PRIOR TO THE RETRACTION OF THE DELIVERY SYSTEM OUTER SHEATH. SIGNIFICANT TENSION AND MANIPULATION OF THE DELIVERY SYSTEM WAS REQUIRED TO UN-SHEATH AND POSITION THE STENT GRAFT. DURING POLYMER FILL, THE FILL POLYMER SYRINGE EMPTIED COMPLETELY AND THE PATIENT EXPERIENCED TRANSIENT HYPOTENSION WHICH WAS SUCCESSFULLY TREATED IN ACCORDANCE WITH THE PRODUCT IFU. THE STENT GRAFT DID NOT FILL WITH POLYMER; CONTRAST WAS INJECTED INTO THE LUMEN OF THE STENT GRAFT TO VISUALIZE THE IPSILATERAL AND CONTRALATERAL LEGS OF THE AORTIC BODY GRAFT TO FACILITATE ACCURATE PLACEMENT OF THE ILIAC LIMBS. THE ANEURYSM WAS SUCCESSFULLY EXCLUDED AND THERE WERE NO ADDITIONAL PATIENT SEQUELAE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 618814 | OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM | ENDOVASCULAR GRAFT, ABDOMINAL AORTIC ANEURYSM | MIH | TRIVASCULAR, INC. | TV-AB2980-D | FS042914-35 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention |