FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 4140874
·
Received October 3, 2014
Report
- Report Number
- 3004753838-2014-25582
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- September 8, 2014
- Report Date
- September 8, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Additional Manufacturer Narrative · 1
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE DEVICE FAILED, CONFIRMING THE REPORTED EVENT OF NO VIBRATION. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE VIBRATE MOTOR.
Description of Event or Problem · 1
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM NO VIBRATION ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, DURING THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT TO TEST THE ALERT FUNCTIONALITY AND REPORTED DEVICE DID NOT VIBRATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 620338 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649-BLU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR |