FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4140856 · Received October 3, 2014

Report

Report Number
2032227-2014-33195
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
September 6, 2014
Report Date
September 6, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACE, MINOR SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP, AND CRACKED BELT CLIP SLOT. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD A BROKEN CLIP SLOT. THE CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620308 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWL

Patients

Seq Age Sex Outcome Treatment
1