FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 4140772
·
Received October 3, 2014
Report
- Report Number
- 1416980-2014-34420
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- September 10, 2014
- Report Date
- September 10, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915523
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION, FUNCTIONAL TESTING, AND A REVIEW OF THE ALARM LOG WERE PERFORMED. FUNCTIONAL TESTING REVEALED THAT THE DEVICE HAD INOPERATIVE FORCE SENSING RESISTORS (FSRS) IN CHANNEL A. THE CAUSE OF THE CONDITION WAS NOT DETERMINED. TO CORRECT THE CONDITION, THE FSRS WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT CHANNEL A OF A FLOGARD INFUSION PUMP WAS OUT OF SERVICE. IT WAS NOT SPECIFIED WHEN IN THE PROCESS THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.