COGNIS
Report
- Report Number
- 2124215-2014-17271
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- September 12, 2014
- Report Date
- October 15, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-0025-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
(B)(4). THE LOCAL AREA SALES REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.
(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. REVIEW OF THE DEVICE MEMORY INDICATED THAT A LOW VOLTAGE ALERT, CODE 1003, WAS RECORDED. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE'S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE'S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE'S BATTERY FASTER THAN NORMAL. BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.
--
ADDITIONAL INFORMATION WAS RECEIVED THAT THE DEVICE WAS EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. DISK ANALYSIS SHOWED THE BATTERY APPEARS TO BE DEPLETING MORE QUICKLY THAN EXPECTED. DEVICE REPLACEMENT WAS RECOMMENDED. DEVICE REPLACEMENT WILL BE PLANNED FOR AN UNKNOWN DATE IN THE FUTURE. AVAILABLE INFORMATION INDICATES THAT THIS DEVICE REMAINS IMPLANTED AND IN SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 618403 | COGNIS | IMPLANTABLE CHF PULSE GENERATOR | LWS | GUIDANT CRM CLONMEL IRELAND | N119 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | 4591| 4469| N119| 0184 |