INSIGNIA
Report
- Report Number
- 2124215-2014-17298
- Event Type
- Malfunction
- Date Received
- October 3, 2014
- Date of Event
- September 12, 2014
- Report Date
- September 29, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- PMA / PMN Number
- D970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
(B)(4). ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT NOISE WAS PRESENT ON THE DEVICE AND RIGHT ATRIAL (RA) LEAD. ADDITIONALLY, RA PACING IMPEDANCE MEASUREMENTS WERE GREATER THAN 2,000 OHMS IN BIPOLAR CONFIGURATION. IN UNIPOLAR CONFIGURATION, PACING IMPEDANCE MEASUREMENTS WERE WITHIN ACCEPTABLE LIMITS. A REVISION PROCEDURE WAS PERFORMED AND THE DEVICE WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE RA LEAD REMAINS ACTIVELY IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED DURING THE PROCEDURE. THE DEVICE WAS RETURNED TO ALLOW FOR RELIABILITY ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 618857 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | 1298 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | K063| 4087| 4086| 1298 |