COGNIS
Report
- Report Number
- 2124215-2014-17247
- Event Type
- Injury
- Date Received
- October 3, 2014
- Date of Event
- September 15, 2014
- Report Date
- October 15, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-0087-2015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. REVIEW OF THE DEVICE MEMORY INDICATED THAT A LOW VOLTAGE ALERT, CODE 1003, WAS RECORDED. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH THE DEVICE'S ABILITY TO DELIVER THERAPY WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE'S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE'S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE'S BATTERY FASTER THAN NORMAL. BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.
(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
(B)(4).
ADDITIONAL INFORMATION BECAME AVAILABLE THAT THIS DEVICE WAS EXPLANTED AND REPLACED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE RECENTLY DISPLAYED A LOW VOLTAGE ALERT CODE OF 1003. BOSTON SCIENTIFIC ENGINEERS WERE CONSULTED AND STATED THAT THE CODE 1003 IS DUE TO LOW VOLTAGES BEING DECLARED DUE TO THREE DAILY VOLTAGES BEING BELOW THE THRESHOLD OF 3.025 VOLTS. THE DEVICE HARDWARE IS NOT DETECTING THE LOSS OF BATTERY ENERGY AND NOT REFLECTING THE DEPLETION CONDITIONS, AND THUS, IS INACCURATE WHICH IS THE REASON FOR THE LOW VOLTAGE ALERT. SINCE THIS DEVICE IS MALFUNCTIONING, IT IS RECOMMENDED TO BE REPLACED IMMEDIATELY. THE DEVICE REMAINS IMPLANTED AT THIS TIME. TO DATE, NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 619325 | COGNIS | IMPLANTABLE CHF PULSE GENERATOR | LWS | GUIDANT CRM CLONMEL IRELAND | N119 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Hospitalization| L| R | 0125| 1851| MISMATCH| N119| T125| 4592 |