FDA Adverse Event Injury Summary report: N

COGNIS

MDR report key: 4140469 · Received October 3, 2014

Report

Report Number
2124215-2014-17328
Event Type
Injury
Date Received
October 3, 2014
Date of Event
September 15, 2014
Report Date
October 17, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
PMA / PMN Number
P010012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND LEFT VENTRICULAR (LV) LEAD EXHIBITED NO PACING OR SENSING AND HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS GREATER THAN 2,000 OHMS. AN INVASIVE PROCEDURE WAS PERFORMED. THE DEVICE AND LEAD WERE EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
618818 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND N118

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| L| R 0185| N161| 4470| N118| 4518| 4592