FDA Adverse Event Injury Summary report: N

OMNIFIT M/S PSL SHELL 48MM

MDR report key: 4140334 · Received October 3, 2014

Report

Report Number
0002249697-2014-03732
Event Type
Injury
Date Received
October 3, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JDI
PMA / PMN Number
K903636
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. ADDITIONAL INFORMATION WAS REQUESTED AND IF IT BECOMES AVAILABLE WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

VISUAL INSPECTION: INSPECTION OF THE PROVIDED PHOTO REGARDING THE SHELL CAN CONFIRM WEAR ON THE RIM OF THE SHELL LIKELY DUE TO DISLOCATION. DURING THE INVESTIGATION IT WAS DETERMINED THAT THE SHELL HAD LOOSENED. THE EVENT WAS CONFIRMED BUT THE ROOT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED DUE TO THE MINIMAL INFORMATION RECEIVED. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME AS NO DEVICES AND INSUFFICIENT INFORMATION WAS RECEIVED BY STRYKER ORTHOPAEDICS. IF DEVICES AND / OR ADDITIONAL INFORMATION BECOME AVAILABLE, THIS INVESTIGATION WILL BE REOPENED.

Additional Manufacturer Narrative · 1

DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: THERE HAVE BEEN NO OTHER EVENTS REPORTED FOR THE REPORTED MANUFACTURING LOT X-RAYS CONFIRM DISLOCATED HIP WITH LOOSE ACETABULAR. NEED CLINICAL HISTORY, OPERATIVE REPORTS AND PROGRESS NOTES. DURING THE INVESTIGATION IT WAS DETERMINED THAT THE SHELL HAD LOOSENED. THE EVENT COULD NOT BE CONFIRMED NOR THE ROOT CAUSE OF THE REPORTED EVENT DETERMINED DUE TO THE MINIMAL INFORMATION RECEIVED. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME AS NO DEVICES AND INSUFFICIENT INFORMATION WAS RECEIVED BY STRYKER ORTHOPAEDICS. IF DEVICES AND / OR ADDITIONAL INFORMATION BECOME AVAILABLE, THIS INVESTIGATION WILL BE REOPENED.

Description of Event or Problem · 1

UPDATE: AS PER X-RAY IMAGE, THE SHELL DISASSOCIATED FROM THE STEM.

Description of Event or Problem · 1

UPDATE: AS PER X-RAY IMAGE, THE SHELL DISASSOCIATED FROM THE STEM.

Description of Event or Problem · 1

REVISION OF TOTAL HIP REPLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
619133 OMNIFIT M/S PSL SHELL 48MM IMPLANT JDI STRYKER ORTHOPAEDICS-MAHWAH ML3108

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention